Fault analysis and treatment method of Wanhao image measuring instrument can't capture image
Wanhao image measuring instrument has two kinds of failures will cause the instrument cannot be used normally in normal use. One is not the image, there is no data two. The two kinds of fault in many cases not instrument hardware failure. But the instrument set out the problem, we carefully analyze, patient treatment, generally can solve their own.
Now to analyze fault. No image:
Analysis.
No image generally indicates a link imaging system problem. Imaging system which consists of several parts? Zoom lens, CCD, CCD power supply, video cable, video card, video card driver. When there is no image, we generally to this pathway one one inspection, such as video cable is plugged. Instrument if there is power etc..
After some analysis, fault part is found, but CCD is not processed by the client to the. Because it relates to the scene debugging and CCD settings. Customers can on the video line, such as normal examination in the component video card. In most cases the problem in these places.
Fault handling:
We first check the video cable is not plugged; using a multimeter to check whether there is open circuit line intermediate; check the video card is inserted firmly. In the temperature changes of the season of PCI socket easily bad contact. General re-inserted video card can solve the problem.
Be careful.
We deal in Wanhao image measuring instrument is not the problem of image, the customer to solve their own problems do not disassemble their repair, so that the expansion of the fault, increased repair costs. If you can't solve, can ask a professional repair personnel to consult or please repair personnel to repair.